How to startup the processor
Konica Minolta User Manuals

Nov 07, 2016 The blots were treated with the ECL kit according to the user manual, developed on X-ray film and finally observed using Konica Minolta SRX-101A.
- Turn power on (right side)
- Close all valves (bottom front, see image)
- Hit Run (front, top right)
- Wait for ready light (about 30 minutes)
Before running a film
- All valves must be closed
- Ready light is on
- Do not hit funnel button beside run, drains all chemicals
- Check developer and fixer supply tanks
- Make sure silver trap in the sink is connected (see image)

How to run a film
- No light in room when film is exposed
- Open developer lid
- Align film on either edge to ensure film runs straight
- Developer will automatically feed
- Close lid, you can turn on lights at this time
- Developer beeps after every film is complete
- Water and chemical will drain into sink drain
How to shutdown the processor
- Turn off power
- Open wash valve to drain water (front left)
- DO NOT open fix or dev valves
Konica Minolta Bizhub User Guide
How to mix chemicals
Konica Minolta User Guides
- When chemicals volumes are below 5 Gal, a new mix is needed
- One bottle of developer + 19 gallons of tap water (fill to 24 Gal line if you started at 5 Gal line)
- Fixer has 2 parts, both bottles + 19 gallons of tap water
- DO NOT get fixer in the developer, if this happens throw out entire batch
- DO NOT mix the stirrers